Filter and collection bottle
The hydrophobic filter retains water and aqueous solu-
NOTICE
tions also in the form of aerosols. When using solvents or
due to evaporation of water, formation of condensate at
the outlet of the BVC is possible.
+ Important: Comply with applicable regulations when
Filter
IN
sartorius stedim
disposing of solvents.
+ Attention: Do not remove connections in case of liq-
uid in the tube. Do not touch contaminated parts when
removing the screw cap. Risk of infection! Wear appro-
priate protective clothing, e.g., gloves.
Change filter if necessary in case of contami-
nation, weak aspiration or no aspiration.
+ Use only the original spare part (see section
"Accessories - Spare parts") and sterilize
and dispose of the filter according to appro-
priate regulations.
Note: After disinfection with alcohol, allow the
filter to dry completely because the filter may
become clogged by alcohol.
page 32 of 76