Adjustment of the BVC level sensor to
work properly with a new bottle, or to
correct false alarms
Pertains only to the BVC professional / BVC professional G
NOTICE
At the factory, the level sensor is adjusted either to the 4l polypropylene
bottle or to the coated 2l glass bottle (G version) at standard conditions.
It may be necessary to adjust the sensor to compensate for changes in
bottle material, bottle size or changed ambient conditions.
NOTICE
We also recommend emptying the bottle and adjusting the BVC profes-
sional if there is a false alarm indicating "liquid level" even though the
bottle is less than 3/4 full.
➨ If there is a false alarm in spite of an adjustment, repeat the adjust-
ment.
➨ If there is a false alarm in spite of repeated adjustment, please call
+49 9342 808-5500 (after sales service), or your local VACUUBRAND
sales office.
NOTICE
Use only the original VACUUBRAND BVC bottle or the original spare bot-
tle (see "Accessories - Spare parts").
Attention: When performing an adjustment, the bottle has to be clean
and free of incrustations or other deposits, see section "Cleaning".
Perform an adjustment only with a clean, empty and dry bottle.
page 40 of 76