Cleaning and decontaminating

Over time, discoloration and material changes (e.g., re-
siliency, elasticity / tightness, cracking) due to repeated
steam sterilizations / autoclavings and / or chemical disin-
fections may occur.
Check all parts regularly. Replace defective parts.

Autoclaving

IN
sartorius stedim
121°C
2 bar absolute
max.
80%
121°C
2 bar
absolute
max. 20 x
page 36 of 76
Autoclaving
The collection bottle with bottle head and screw
cap, the quick coupling and the filter are des-
ignated for steam sterilization at 121°C and 2
bar absolute (1 bar overpressure). Time of ex-
posure according to DIN 58946 t
NOTICE
Prior to autoclaving loosen or remove the
bottle head from the bottle.
It is the user´s responsibility to ensure effective
autoclaving.
NOTICE
Autoclaving
Maximum liquid level in collection bottle:
80%. When working with low boiling liquids or
liquids that tend to foam, the maximum liquid
level may be less.
The number of autoclaving cycles can be
marked on the plastic disc (Memory Disc) of the
filter (max. 20 autoclaving cycles according to
the manufacturer instruction).
= 20 minutes.