Operation Bvc Control / Bvc Control G - vacuubrand BVC control Instruction Manual

Biochem-vacuucenter
Hide thumbs Also See for BVC control:
Table of Contents

Advertisement

Operation BVC control / BVC control G

The unit has a touch panel so the keys need only be touched gently.
Do not press the keys!
The keys "+" and "-" have to be touched > 0.25 seconds to be actuated.
A successful action is confirmed by a blip and the flashing of LEDs.
1
-
Use the keys to set suction power of the system.
The suction power can be set linearly in a range from 150 mbar (1
LED flashes) to 850 mbar (8 LEDs flash) underpressure (relative to
+
atmospheric pressure).
page 24 of 76
Position
2
3
A flashing LED indicates that the actual avail-
able suction power differs from the preset suc-
tion power.
Attention: If the collection bottle is under
vacuum and the vacuum demand is reduced,
the existing underpressure inside the bottle
remains until the vacuum is reduced (pres-
sure rises) by further aspiration or venting
with the VHC
pro
Venting the BVC: Press the lever of the VHC
or set the VHC
Designation
1
Key to reduce suction power
2
Display suction power
Key to increase suction
3
.
to continuos aspiration.
pro
power
pro

Advertisement

Table of Contents
loading

This manual is also suitable for:

Bvc control gBvc professionalBvc professional g

Table of Contents