Aa1 - ICP DAS USA Network Device I-7011 User Manual

B&b electronics network device user manual
Table of Contents

Advertisement

2.6 $AA1

Description : Perform Zero Calibration
Syntax : $AA1[CHK](cr)
$
delimiter character
AA
address of setting module (00 to FF)
1
command for performing zero calibration
Response :
Invalid Command :
Syntax error or communication error may get no re-
sponse.
!
delimiter for valid command
?
delimiter for invalid command or the calibration is not
enabled
AA
address of response module(00 to FF)
Example :
Command : $011
Perform address 01 zero calibration, return success.
Command : $021
Perform address 02 zero calibration, return the calibration is
not enabled before perform calibration command.
Related Command :
Sec.2.5 $AA0, Sec.2.20 ~AAEV
Related Topics :
Sec.1.9 Calibration
Rev:B1.0
Valid Command :
I-7011, I-7018 User Manual
!AA[CHK](cr)
?AA[CHK](cr)
Receive : !01
Receive : ?02
27

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

I-7011dI-7011pI-7011pdI-7018I-7018p

Table of Contents