2 Installation
Installation
The safety of any system incorporating the equipment is the responsibility of the assembler
of the system.
The instrument requires a positive pressure supply, instruments operating in an absolute
range or negative pressure range require a vacuum supply.
A vacuum supply should be used for a fast response for instruments operating near
atmospheric pressure.
For dual channel operation two independent pressure and vacuum supplies can be used.
Note: When using two pressure modules
make sure that:
•
The module with the highest pressure rating is fitted to the right hand side
Module 1 position as viewed from the rear of the product refer to
•
If two modules have the same pressure rating, make sure that the module with
the higher serial number is fitted to the right hand side Module 1 position as
viewed from the rear of the product.
Note: All pneumatic connections must comply with the Pressure Equipment Directive (PED)
or other regional pressure standards.
Note: When connecting the output ports of two pressure modules together make sure both
are either:
•
below 70 bar
OR
•
between 100 to 210 bar.
To prevent over-pressurisation of pneumatic parts and maintain compliance with the PED do
not mix categories.
K0443 Revision A
Figure 2-2, Pressure module rear view
(Ref: Section 4.9, Pressure module replacement)
2 - 4
(Ref: Fig 2-2)
[EN] English