Tare enabled.
3
Press to switch between controlled pressure and measured pressure.
4
Nudge up.
5
Set-point, press and the display changes to numeric keys.
6
Nudge down.
7
Status area shows effort meter set in global set-up, press to enter control set-up.
Notes:
Active mode - control active, except in measure mode.
Passive mode - when the controller achieves inlimits condition, measure mode is
automatically selected.
Gauge mode - when the controller achieves zero gauge inlimits condition, measure mode is
automatically selected and the zero valve opens.

Controlling to a new set-point

1.
To change the set-point value, touch the set-point area of the screen and the display
shows the numeric keys.
2.
Set the new set-point value.
3.
If necessary, use the
4.
To save the new set-point, touch the set-point area of the screen. This returns the
display to the measured pressure screen and showing the new set-point.
5.
Touch the escape
6.
To control pressure to the new set-point value press the
7.
The display shows the pressure value changing as the instrument controls to the new
set-point, at the set rate of change.
Note: When changing from measured to controlled mode the displayed pressure digits
change from black (measured pressure) to blue (controlled pressure out of limits) to
green (controlled pressure in-limits).
8.
If enabled, the effort meter shows the effort the controller exerts to achieve the set-
[EN] English
key to remove the last digit in the set-point value display.
to leave the numeric setting unchanged.
3 - 7
key.
Control
K0443 Revision A