Page 3
If repairs or modifications are made by personnel not authorized by Olympus, the warranty is void, and Olympus shall not be liable for damage that occurs to or as a result of use of the modified product. Applicable Unit •...
Foot switch port for MAJ-1258 ..............1-14 Bipoler foot switch port for MAJ-1259............1-14 System port ....................1-15 System Diagram ..................1-16 Endoscopic Treatment with UES-40 - System Diagram ......1-16 Electrosurgical Treatment with UES-40 - System Diagram .......1-18 Chapter 2: Troubleshooting........... 2-1 Troubleshooting ..................2-2 Troubleshooting list ..................2-3...
Page 5
UES-40 Chapter 4: Inspection ............. 4-1 Preparation ....................4-2 Inspection Procedure................4-2 Chapter 5: Parts List............... 5-1 Exploded Parts Diagram................5-2 Parts List....................5-6 ISSUE4 CONTENTS...
Chapter 1: Product Specification Chapter 1:Product Specification Outline ..............1-2 Specifications ............1-2 Operating Requirements.......... 1-8 Name and Function of Each Part......1-10 Ports............... 1-14 System Diagram ............ 1-16 ISSUE4 Product Specification...
(1) Expanding the functions available with the UES-30 The UES-40 also offers an output system for use in physiological saline (TURis). TURis frees surgeons from the need for a patient plate, and offers improved safety over con- ventional TUR is.
Page 9
UES-40 • Saline system Cutting Mode: 2 options (Pure, Blend) Coagulation Mode: 2 options (Coag 1, Coag 2) 2-1-3 Basic frequencies 350 kHz, 1 MHz (when using Spray) 2-1-4 Output characteristics • Bipolar system Bipolar: Maximum setting Pure Soft 1, Soft 2...
Page 10
Control switches l Foot switch for UES-40 (MAJ-1258) Controls the high-frequency output of monopolar and bipolar instruments (ON/OFF) l Bipolar foot switch for UES-40 (MAJ-1259) Controls the high-frequency output of bipolar instruments (ON/OFF) l Handpiece output switch Controls the high-frequency output of the handpiece (ON/OFF)
Output System Output terminal Control system Remarks option Bipolar output can be Foot switch for UES-40 Bipolar 1 port Coag, Cut selected using the Foot- (MAJ-1258) switch select switch. Bipolar Bipolar foot switch for...
UES-40 2-2-2 High-frequency-output monitor Detects whether the output value is correct for the output setting value. If problems are detected, the following operations are performed. • Error code is displayed on output indicator. • Audible warning is issued. • High-frequency output is stopped.
Class-I equipment (three-prong plug for power supply) l Level of electrical-shock protection CF-type (low-frequency leakage current of 10µþA or less) l Level of explosion protection Never use the UES-40 where flammable vapors occur. l Others Defibrillator protection Laws, regulations, and standards...
(5) Do not attach the patient plate for use near metallic implants. (6) Do not wind UES-40 cords into loops or allow them to touch the patient or metallic parts of the surgical table. (7) Do not bundle UES-40 cords with cords of other medical equipment during use.
Page 15
UES-40 Bipolar: Cut, Soft 1, Soft 2 Soft1, Soft2 Output setting (W) Figure 1-1 Open voltage characteristics (Bipolar: Pure, Soft 1, Soft 2) Bipolar: Hard Hard Output setting (W) Figure 1-2 Open voltage characteristics (Bipolar: Hard) Monopolar 3500 3000 Pure/Urology...
UES-40 4. Name and Function of Each Part Front panel (for 120 V and 220 V) 4-1-1 Switches and ports (4) Bipolar coagulation mode (7) Monopolar/saline cut mode select switch select switch (1) Memory switch (5) Saline switch (8) Monopolar/saline coagulation...
Page 17
UES-40 (6) Foot-switch select switch This switch is used to switch between the saline/monopolar mode and the bipolar output mode. (7) Monopolar/saline cut mode select switches These switches are used to select the monopolar and saline cut mode (8) Monopolar/saline coagulation mode select switches These switches are used to select the monopolar and saline coagulation mode.
Page 18
This indicator displays the value specified using the output control switches. (21) Foot-switch setting indicator This indicator displays the available output from the foot switch for the UES-40 (MAJ-1258). (22) Monopolar/saline cut output setting indicator This indicator displays the value specified using the output control switches.
UES-40 (25) Monopolar/saline cut mode indicators The lighted indicator corresponds to the cutting mode selected using the switches. (26) Saline mode indicator This indicator lights up when saline mode is selected using the switch. (27) Bipolar coagulation mode indicators The lighted indicator corresponds to the coagulation mode selected using the switches.
UES-40 (3) Foot switch port for MAJ-1258 This port is for connecting the standard accessory foot switch for UES-40. (4) Foot switch port for MAJ-1259 This port is for connecting an optional bipolar foot switch for UES-40. (5) Circuit breakers These breakers are used in the event of mechanical problems to prevent electrical surges within the UES-40.
UES-40 System port Pin No. Signal Pin No. Signal TxD1 RxD1 RTS1 CTS1 DSR1 DTR1 SMKEVAC1 13 12 11 10 9 25 24 23 22 21 20 19 18 17 16 15 14 As viewed from the connection side ISSUE4...
Disposable P-cord (MAJ-814) Some products may not be available in some regions. Please contact the nearest Olympus office for details Specifications, design and accessories are subject to change without any notice or obligation on the part of the manufacturer. Product Specification...
Page 23
UES-40 l 220 V specification Videoscope without the S-cord connector mount Bipolar foot switch for UES-40 (MAJ-1259) Fiberscope without the S-cord connector mount Foot switch (MAJ-1258) Interactive cable (MAJ-877) High flow insufflation unit (UHI-2) UES-40 High flow insufflation unit (UHI-3)
UES-40 Electrosurgical Treatment with UES-40 - System Diagram l 120 V specification Resectoscope for saline Saline electrodes Sheath (WA22301D), (WA22302D) Saline working elements (WA22305D), (WA22306D) (WA22060B) (WA22331D), (WA22332D) (WA22061B) (WA22351A), (WA22351C) (WA22355A), (WA22355C) Working elements (A22060A), (A22061A) (A20063A) Sheath Electrode...
Page 25
UES-40 A adapter (MAJ-1) is not available in some countries. **: A adapter 2 (MAJ-619) is required when the A-Cord marked (**) is used. *: A adapter 3 (MAJ-620) is required when the A-Cord marked (*) is used. (WA00013A) Foot switch...
Page 26
UES-40 l 220 V specification Resectoscope for saline Saline electrodes Sheath (WA22200D) Saline working elements (WA22232D) (WA22060B) (WA22208D) (WA22061B) (WA22209D) (WA22233D) Working elements (A22060A), (A22061A) (A20063A) Sheath Electrode Working elements (A2760), (A2761) Working elements (A2749), (A2750) (A2751), (A2752) (and other cystourethroscopes)
Page 27
UES-40 A adapter (MAJ-1) is not available in some countries. **: A adapter 2 (MAJ-619) is required when the A-Cord marked (**) is used. *: A adapter 3 (MAJ-620) is required when the A-Cord marked (*) is used. (WA00013A) Foot switch...
UES-40 1. Troubleshooting Configuration of the main unit UPHF40CR00U (control board) UPHF40OT00U (oscillation board) UPHF40EX00U (relay board) CONVERTER UPHF40SS00U (monitor board) Troubleshooting ISSUE4...
Page 31
Parts That Might Have Caused the Problem Problem Oscillation Monitor Relay Control Panel Converter Harnesses Other Other than main unit board board board board sheet Power-on Converter fan unresponsive — — — — — Converter – Power switch Power switch —...
Page 32
Parts That Might Have Caused the Problem Problem Oscillation Monitor Relay Control Panel Converter Harnesses Other Other than main unit board board board board sheet There is a No output Converter – Oscillation board (J1) response when Monitor board (J3) – Oscillation board (J7) the output switch is Monitor board (J3) –...
Page 33
Parts That Might Have Caused the Problem Problem Oscillation Monitor Relay Control Panel Converter Harnesses Other Other than main unit board board board board sheet When used with Does not function in a ganged manner • Smoke-controlling cable discon- the UHI, does nected not control •...
Page 34
Parts That Might Have Caused the Problem Problem Oscillation Monitor Relay Control Panel Converter Harnesses Other Other than main unit board board board board sheet (From previous High-frequency current monitor dam- Monitor board (J7) – Relay board (J2) Used where the high-frequency cur- age (Cut 01) rent is prone to leakage.
UES-40 1. Precautions for disassembly and reassembly (1) Keep all components safe, since you will need to reattach them at their original positions and reroute the cables as originally routed. (2) Exercise extreme caution to prevent possible injuries during disassembly.
UES-40 2. Disassembly procedure Tools Name Specifications Remarks Box screwdriver 5 mm between opposite sides Box screwdriver 5.5 mm between opposite sides Box screwdriver 8 mm between opposite sides Wrench 19 mm between opposite sides Neji-lock 1401C Red Phillips screwdriver No.1...
UES-40 Disassembly procedure 2-2-1 UES-40 main-unit disassembly HCBK3x6SA (1) Remove the 15 SCREWS holding the TOP 2 SCREWS CBK3x6SA COVER (HCBK3x6SA: 14 SCREWS; 1 SCREW CBK3x6SA: 1 SCREW) and remove the TOP COVER. TOP COVER Phillips screwdriver (No. 2) Note: During reassembly, make sure no CABLES are caught under the COVER.
Page 41
UES-40 (4) Detach the E23 HARNESS from J7 of UPHF40SS00U (monitor board). E23 HARNESS UPHF40SS00U (monitor board) C6N3SZ 3 NUTS (5) Remove the 3 NUTS holding the FRONT PANEL UNIT (C6N3SZ) and detach the FRONT PANEL UNIT. FRONT PANEL UNIT Box screwdriver (5.5 mm between opposite sides)
Page 42
UES-40 E24F CABLE E12 HARNESS (7) Detach the E12 HARNESS from J3 of UPHF40CR00U UPHF40SS00U (monitor board) and J7 of (control board) UPHF40OT00U (oscillation board). UPHF40OT00U (oscillation board) (8) Detach the E24F CABLE from J1 of UPHF40CR00U (control board) and J16 of UPHF40SS00U (monitor board).
Page 43
UES-40 E12 HARNESS 2 (12) Detach the E12 HARNESS 2 from J6 of UPHF40OT00U (os- cillation board) UPHF40SS00U (monitor board) and J3 of UPHF40OT00U (oscillation board). UPHF40SS00U (monitor board) VOLUME HARNESS SPEAKER HARNESS (13) Detach the VOLUME HARNESS from J9 of...
Page 44
UES-40 4 SPACERS UPHF40SS00U (15) Remove the 4 SPACERS between (monitor board) UPHF40CR00U (control board) and CCUK3x6SZ 2 SCREWS UPHF40SS00U (monitor board). Box screwdriver (5.5 mm between opposite sides) (16) Remove the 2 SCREWS (CCUK3x6SZ) hold- ing UPHF40SS00U (monitor board), and remove UPHF40SS00U (monitor board).
Page 45
UES-40 CCUK3x6SZ 6 SCREWS (18) Remove the 6 SCREWS (CCUK3x6SZ) hold- ing UPHF40OT00U (oscillation board), and UPHF40OT00U detach UPHF40OT00U (oscillation board). (oscillation board) Phillips screwdriver (No. 2) CONVERTER C6N3SZ 4 NUTS (19) Remove the 4 NUTS (C6N3SZ) holding the CONVERTER.
Page 46
UES-40 2-2-2 REAR PANEL UNIT disassembly (1) Remove the CAP. (2) Remove the single attached SCREW (for the VOLUME HARNESS) and remove the PEAR PANEL KNOB. Attached Phillips screwdriver (No. 1) SCREW Neji-lock (during reassembly) Note: During reassembly, pay attention to the KNOB following items.
Page 47
UES-40 (3) Remove the 2 NUTS (C6N3SZ) and 2 C6N3SZ WASHERS (HWB3SA) holding the EARTH 2 NUTS CABLE, and remove the EARTH CABLE. Box screwdriver (5.5 mm between opposite sides) HWB3SA 2 NUTS EARTH CABLE REAR PANEL SYSTEM HARNESS C6N3SZ...
Page 48
UES-40 Attached (6) Remove the attached NUT and 2 WASHERS Attached CUTB3x8SZ WASHER 4 SCREWS holding the VOLUME HARNESS, and remove the VOLUME HARNESS. Box screwdriver (8 mm between opposite sides) Neji-lock (during reassembly) Note: During reassembly, apply Neji-lock to the HARNESS threads.
Page 49
UES-40 2-2-3 FRONT PANEL UNIT disassembly FRONT CHASSIS (1) Remove the 7 SCREWS (CUKSK3x6SZ) CUKSK3x6SZ holding the FRONT CHASSIS, and remove 7 SCREWS the FRONT CHASSIS. Phillips screwdriver (No. 2) Note: During reassembly, thread each of the 3 HARNESSES through the FRONT CHASSIS holes.
Page 50
UES-40 NN2.5SA 2 NUTS (4) Remove the 2 NUTS (NN2.5SA) and 4 SW2.5SA 2 WASHERS WASHERS (2 WASHERS each of SW2.5SA and KNW2.5SA) holding UPHF40EX00U KNW2.5SA 2 WASHERS (relay board). UPHF40EX00U Box screwdriver (5 mm between opposite sides) (relay board)
Page 51
UES-40 SPACER 23.7 (7) Remove the 2 SCREWS (CUKSK3x6SZ), and then remove the SPACER 23.7, CON- TACT SPRING, SPACER 25, and CONTACT CUKSK3x6SZ PLATE (in that order). CONTACT SPRING Box screwdriver (5.5 mm between opposite sides) Phillips screwdriver (No. 2)
UES-40 3. Reassembly procedure Tools Please refer to “2-1 Tools.” Reassembly procedure To perform reassembly, reverse the disassembly procedure, referring to “2-2 Disassembly procedure.” No adjustments such as positioning are required during reassembly. Disassembly and Reassembly Procedure 3-16 ISSUE4...
– DNI NEVADA INC. 2. Inspection Procedure Connect the Electrosurgical Analyzer to the UES-40. (Refer to the Electrosurgical Analyzer instruction manual for details.) Refer to the following table to check output values: Table 4-1 Setting and output of each mode Output mode Load resistance [¾]...
UES-40 1. Exploded Parts Diagram HCBK3x6SA HCBK3x6SA CBK3x6SA GL465300 HCBK3x6SA HCBK3x6SA GT011600 EXPLODED PARTS DIAGRAM MODEL UNIT FR.1590 UES-40 TOP COVER FIG.1/4 OLYMPUS MEDICAL SYSTEMS CORP. (C) 2004 Olympus Medical Systems Corporation. All rights reserved. Parts List ISSUE4...
Page 57
DY266000 GC475800 GT017300 DO078500 GT013100 C6N3SZ GT203000(220V) HCBK3x6SA C6N3SZ HCBK3x6SA GC980900 GC981300 EXPLODED PARTS DIAGRAM UNIT FR.1590 MODEL FIG.2/4 FRONT PANEL , CONVERTER UES-40 OLYMPUS MEDICAL SYSTEMS CORP. (C) 2004 Olympus Medical Systems Corporation. All rights reserved. Parts List ISSUE4...
Page 58
C6N3SZ GL356300 GL357300 CCUK3x6SZ GN380600 (English) GN380700 (Symbol) Upgrade kit EXPLODED PARTS DIAGRAM GL335500(120V) MODEL UNIT FR.1590 GU197300(220V) UES-40 PCB , REAR PANEL FIG.3/4 OLYMPUS MEDICAL SYSTEMS CORP. (C) 2004 Olympus Medical Systems Corporation. All rights reserved. Parts List ISSUE4...
Page 59
GL357300 GL356000 SYSTEM FRONT PANEL SPEAKER VOLUME (For MAJ-1258) (For MAJ-1259) GL355700 GL355500 GL355900 GL355600 EXPLODED PARTS DIAGRAM MODEL UNIT FR.1590 UES-40 CABLE FIG.4/4 OLYMPUS MEDICAL SYSTEMS CORP. (C) 2004 Olympus Medical Systems Corporation. All rights reserved. Parts List ISSUE4...
Page 60
UES-40 2. Parts List PARTS No. INDEX PARTS NAME E. SPECIFICATION REMARK DO078500 2-D2 BINDING WIRE DY266000 2-C2 SWITCH DZ253200 3-A2 FERRITE CORE GC475800 2-A2 BUSH 4 GC688300 2-D2 SWITCH SPRING GC688400 2-D2 SWITCH COLLAR GC688500 2-D1 MAIN SWITCH GC980700...
Page 61
UES-40 PARTS No. INDEX PARTS NAME E. SPECIFICATION REMARK GN380600 3-D3 VER2 SET-E Ver.2 English GN380700 3-D3 VER2 SET-S Ver.2 Symbol GT011600 1-C2 MYLAR T GT011800 3-A1 REAR PANEL E 120V GT011900 3-A1 REAR PANEL S 220V GT012500 2-C1 CONTACT PLATE...
Page 64
2951 Ishikawa-cho, Hachioji-shi, Tokyo Japan Service Engineering Medical Service Engineering Depertment Production and Service Division PRINTED IN JAPAN FIRST EDITION : 05.2004 W102B...