6.3

Ultrasonic cleaning

VISERA RHINO-LARYNGO VIDEOSCOPE ENF-V2
Chapter 6 Recommended Reprocessing Methods and Chemical Agents
Use a medical-grade, low-foaming detergent for ultrasonic cleaning (neutral pH
detergent or enzymatic detergent) and follow the manufacturer's dilution and
temperature recommendations. Do not reuse detergent solutions.
Use a medical-grade ultrasonic cleaner with a sufficient size and depth to allow
complete immersion of all components of the instrument.
Operate the ultrasonic cleaner for at least 5 minutes (max. 15 minutes) at a
frequency of 38 – 47 kHz.
Switch off the cleaner's heating system. Components must be cleaned at room
temperature.
Vapor generated by the ultrasonic cleaning device may
contain infectious agents such as aerosol. Always use a face
screen and air exhaust during ultrasonic cleaning.
Excessive detergent foaming can prevent fluid from
adequately contacting internal lumens (e.g. the inside of the
insertion section).
Use only those detergents which feature validated processes
in accordance with national and local regulations and/or
guidelines. There is an infection risk when using insufficient
detergents.
Use only solutions whose compatibility for cleaning
endoscopic instruments is certified by manufacturers, and
which are approved by a competent authority. Incompatible
detergents may considerably damage Olympus endoscopic
instruments.
For details on the operation of the ultrasonic cleaner, refer to
its instruction manual.
39