Errors - Ametek JOFRA ASM-801 Reference Manual

Advanced signal multi-scanner
Table of Contents

Advertisement

8.0

Errors

Warning
During set-up, make sure that there is correlation between
the software set-up, and the actual distribution of the sensors
connected to the multi-scanner. The software cannot detect
misplacements of sensors.
If the power is lost to the entire test set-up or part of it, it is highly
recommended to restart the calibration.
GENERAL ERRORS
Error description
LED channel 1 does not flash :
Faulty readout :
No contact with the multi-
scanner :
125614 04
Solution
Check the power supply
Switch the multi-scanner off and on
again
If the error still occurs, return the
multi-scanner to the manufacturer for
service
Check that the connected sensors are
correctly connected and that they are
properly polarised
Check that the software is set up
correctly
Switch the multi-scanner off and on
again
If the error still occurs, return the
multi-scanner to the manufacturer for
service
Check the power supply
Check the RS232 connection. Use
only the supplied cable
Switch the multi-scanner off and on
again - and set up the software again
If the error still occurs, return the
multi-scanner to the manufacturer for
service