Type a unique name for the measurements. Click As found
and ensure that the date is correct. This information will
appear in a measurement report displayed on screen.
Click OK to view the measurement report.
Note...
These results are not saved electronically. If you require a
record of these measurements, click Print while it is
displayed.
When you are finished viewing or printing the report, click
Close. This returns you to the first dialogue in the
Temperature adjustment option.
If you do not require any further measurements, click Done
to return to the Main Menu.
9.2.6

To adjust and calibrate (as left) the ASM-80x

Note...
If you select Cold Junction compensation, either you must
also select the mV (TC) Input or be sure that the mV(TC)
range is calibrated.
Click the radio button Adjust/Calibrate.
Use the check box list to select the inputs to be measured.
If you have selected more than one check box, the
measurements occur in the same order as they appear in the
check box list.