To Calibrate (As Found), Adjust And Calibrate (As Left) The Asm-80X - Ametek JOFRA ASM-801 Reference Manual

Advanced signal multi-scanner
Table of Contents

Advertisement

When you are finished viewing or printing the report, click
Close. This returns you to the first dialogue in the
Temperature adjustment option.
If you do not require any further measurements, click Done
to return to the Main Menu.
9.2.7
To calibrate (as found), adjust and calibrate (as left) the
ASM-80x
Note...
If you select Cold Junction compensation, either you must
also select the mV (TC) Input or be sure that the mV(TC)
range is calibrated.
Click the radio button Measure/Adjust/Calibrate.
Use the check box list to select the inputs to be measured.
If you have selected more than one check box, the
measurements occur in the same order as they appear in
the check box list.
Connect the reference signal source to the appropriate
channel (when prompted).
Click Start and follow the instructions on screen.
Each input has several pre-set input values. Ensure that the
reference signal source is adjusted correctly when
prompted.
For Cold Junction compensation, ensure the corresponding
µV value for the calibrated junction temperature is entered
(0µV @ 0°C).
The software measures the pre-set values for the selected
input, adjusts the settings and then re-calibrates the
calibrator's inputs. All this is done before moving on to the
next selected input.
48
2017-08-11
125614
04