Wai; System_Idn; Sys_Conf - Ametek JOFRA ASM-801 Reference Manual

Advanced signal multi-scanner
Table of Contents

Advertisement

6.3.13

*WAI

SCPI standard command (Wait-to-continue).
Because the system only implements sequential commands this
command has no functionality.
6.3.14

SYSTEM_IDN?

Returns IDN-information for all ASM-units connected in a format like
the *IDN-command. However, it is extended with the box number (fist
field) and calibration date (last field), e.g.:
1,AMETEK,ASM801B,123456-11111,0.01,2005-06-28
2,AMETEK,ASM802A,123456-22222,0.01,2005-06-28
3,AMETEK,ASM803A,123456-33333,0.01,2005-06-28
6.3.15

SYS_CONF

SYS_CONF <Freq.> Hz,<Speed>
<Freq> = 50, 60
<Speed> = 1,2,3,4
Sets up A/D converter speed/filters in all connected units. The speed
parameter is only used by B-models. However, the speed parameter
must always be specified.
Default settings at power up is 50 Hz and speed 1.
6.3.16
SYS_CONF?
Returns the actual SYS_CONF configuration.
Example 50 Hz, 3
If the connected units have different configurations "ERROR" will be
returned.
125614 04
2017-08-11
29

Advertisement

Table of Contents
loading

Table of Contents