6.3.7

*OPC?

SCPI standard command (Operation Complete query).
Because the system only implements sequential commands this
command will immediately output a "1" when the command is
executed.
6.3.8

*RST

SCPI standard command (Reset). In this implementation the
execution of this commands correspond to a power off/on.
6.3.9
*SRE <reg val>
SCPI standard command (Service Request Enable).
Set the Service Request Enable register, which is a 8-bit register, i.e.
a value between 0 and 255 must be specified as argument.
6.3.10

*SRE?

SCPI standard command (Service Request Enable query).
Outputs the value of the Service Request Enable register.
6.3.11

*STB?

SCPI standard command (Status Byte query).
Outputs the value of the Status Byte register. In this implementation
only the ESB bit (Event Summary Bit) is used, i.e. the MAV bit
(Message Available) and RQS/MSS bit is always set to 0.
The ESB bit is set if one of the enabled events in the ESR register
has occurred since last read of the STB register. The ESB bit is
always cleared after a read.
6.3.12

*TST?

SCPI standard command (Self-Test). In this implementation no self-
test is performed when executing this command, i.e. this command
always return OK.