10 Maintenance 
Our  sales  representatives  and  customer  service  are  always  at  your  disposal  for  questions  regarding 
application assistance, calibration, repair, and solutions to specific problems. Please contact your local 
sales representative if you need assistance. In many cases, problems can be solved over the telephone. 
If  you  need  to  return  equipment  for  servicing,  calibration,  or  repair,  please  contact  our  Service 
Department before shipping. Phone numbers are listed at the beginning of this document. 
10.1 Troubleshooting Minor Problems 
Symptom
No output
No power to instrument
Erroneous temperature Faulty sensor cable
Erroneous temperature Field of view obstruction
Erroneous temperature Window lens
Erroneous temperature Wrong slope or emissivity
Temperature fluctuates Wrong signal processing
Marathon Series FA/FR 
Probable Cause
Table 8: Troubleshooting 
Solution
Check the power supply
Verify cable continuity
Remove the obstruction
Clean the lens
Correct the setting
Correct Peak/Valley Hold or Average settings

Maintenance

45