To View A Two-Tone Waveform - Keysight E8267D User Manual

Psg signal generators
Hide thumbs Also See for E8267D:
Table of Contents

Advertisement

Two–Tone Waveform Generator
Creating, Viewing, and Modifying Two–Tone Waveforms
Figure 10-2
To View a Two–Tone Waveform
This procedure describes how to configure the spectrum analyzer to view a
two–tone waveform and its IMD products. Actual key presses will vary,
depending on the model of spectrum analyzer you are using.
1. Preset the spectrum analyzer.
2. Set the carrier frequency to 20 GHz.
3. Set the frequency span to 60 MHz.
4. Set the amplitude for a 10 dB scale with a 4 dBm reference.
5. Adjust the resolution bandwidth to sufficiently reduce the noise floor to
6. Turn on the peak detector.
7. Set the attenuation to 14 dB, so you're not overdriving the input mixer on
You should now see a two–tone waveform with a 20 GHz center carrier
frequency that is similar to the one shown in
products at 10 MHz intervals above and below the generated tones, and a
carrier feedthrough spike at the center frequency with carrier feedthrough
distortion products at 10 MHz intervals above and below the center carrier
frequency.
264
expose the IMD products. A 9.1 kHz setting was used in our example.
the spectrum analyzer.
Figure
10-3. You will also see IMD
Keysight E8357D/67D & E8663D PSG User's Guide