Analog Modulation
Configuring the LF Output (Option UNT)
Internal modulation (Internal Monitor) is the default LF output source.
NOTE
Configuring the Internal Modulation as the LF Output Source
1. Press Preset.
2. Press the FM/ΦM hardkey.
3. Press FM Dev > 75 > kHz.
4. Press FM Rate > 10 > kHz.
5. Press FM Off On.
You have set up the FM signal with a rate of 10 kHz and 75 kHz of deviation.
The FM annunciator is activated indicating that you have enabled frequency
modulation.
Configuring the Low Frequency Output
1. Press the LF Out hardkey.
2. Press LF Out Amplitude > 3 > Vp.
3. Press LF Out Off On.
You have configured the LF output signal for a 3 volt sine wave (default wave
form) output which is frequency modulated using the Internal 1 Monitor
source selection (default source).

To Configure the LF Output with a Function Generator Source

In this example, the function generator is the LF output source.
Configuring the Function Generator as the LF Output Source
1. Press Preset.
2. Press the LF Out hardkey.
3. Press LF Out Source > Function Generator 1.
Configuring the Waveform
1. Press LF Out Waveform > Swept–Sine.
2. Press LF Out Start Freq > 100 > Hz.
3. Press LF Out Stop Freq > 1 > kHz.
4. Press Return > Return.
182
This returns you to the top LF Output menu.