M9383B VXG-m PXIe Microwave Signal Generator
Startup Guide
Step 5: Generate and View an Output Signal
After verifying the operation of the M9383B VXG-m in
measurement.
The following measurement uses a M9383B VXG-m to generate
the 2 GHz signal and Keysight Signal Analyzer to analyze it. You
may use any frequency depending upon the signal analyzer used.
1. Power on the Keysight Signal Analyzer.
2. Connect a high-quality male (2.4 mm) to SMA (male) cable between the RF Out
connector on the M9323A Amplifier and the RF In connector on the Keysight Sig-
nal Analyzer.
3. Torque the connectors to 8 In-lb (0.904 Nm).
4. On the M9383B VXG-m, set the following parameters:
a. Frequency: 2 GHz
b. Power: -10 dBm
c. Select the ALC On checkbox in the RF Output Block.
d. Set the Modulation and RF Out switches to ON.
e. Set the RF Out (All) switch to ON.
Step
4, you are ready to make a
77