Errors.

Table 6-2. Error messages.
Table 6-2 is a list of diagnostic messages that may be displayed on the
Integral Display, on the Model 275 HART Communicator, in AMS or by the
Radar Configuration Tools (RCT) software. Errors normally result in Analog
Output alarm.
Errors are indicated in RCT by the message "Transmitter malfunction":
To see the error message do one of the following:
Click the Read Gauge Status icon
RCT workspace.
1. Open the Advanced section in the RCT workspace Project Bar and
click the Device Commands icon,
or
choose the Device Commands option from the View menu.
2. Open the folder named Diag and double-click the Read Gauge
Status option.
Message
Invalid configuration.
At least one configuration parameter
is outside allowed range.
NOTE: the default values are used
until the problem is solved.
RAM failure was
The transmitter performs an
detected during
immediate reset.
startup test.
FPROM failure was
The transmitter performs an
detected during
immediate reset.
startup test.
Waveform acquisition
This error is probably caused by
failure.
hardware failure.
EEPROM factory
Checksum error in the factory
checksum.
configuration parameters. Can be
caused by power failure during
configuration or by hardware error.
NOTE: the default values are used
until the problem is solved.
EEprom user
Caused by error in the User
checksum error.
Configuration parameters. Can be
caused by power failure during
configuration or by hardware error.
NOTE: the default values are used
until the problem is solved
Software error.
Probe failure.
Probe is not detected.
Rosemount 3300 Series
in the toolbar at the top of the
Description
• Load default database and restart
• Contact Saab Rosemount service
Contact Rosemount service
department.
Contact Rosemount service
department.
Contact Rosemount service
department.
Contact Rosemount service
department.
• Load default database and restart
• Contact Saab Rosemount service
Contact Rosemount service
department.
Check that the probe is correctly
mounted.
Error indication
Action
the transmitter.
department if the problem persists.
the transmitter.
department if the problem persists.