Emerson Rosemount 3300 Series Reference Manual page 36

Guided wave radar level and interface transmitters
Hide thumbs Also See for Rosemount 3300 Series:
Table of Contents

Advertisement

Rosemount 3300 Series
3-14
Coaxial
L>49 inches/
1250 mm
Maximum shortening
23.6 inches/
600 mm
Minimum
15.7 inches/
400 mm
L≤49 inches/
1250 mm
Reference Manual
00809-0100-4811, Rev AB
1. Cut the pipe to the desired
length.
2. Cut the rod inside the pipe.
Make sure that the rod is fixed
while cutting.
NOTE!
Be careful when cutting the inner
rod. Too much force on the inner
rod may cause the spacers
centering the rod to come loose.
3. Update the transmitter
configuration to the new probe
length.
• Pipes longer than 49
inches/1250 mm can be
shortened by as much as 23.6
inches/600 mm.
• Pipes shorter than 49
inches/1250 mm can be cut as
long as the remaining length is
not less than 15.7 inches/400
mm.
July 2003

Advertisement

Table of Contents
loading

Table of Contents