Table 22. Library error codes (continued)
Code (H)
Description
X movement error #1.
During X movement, the target
0080
stop position's origin sensor error
detected. Suspect the X origin
sensor or X motor.
X movement error #2.
During X movement, a motor
0081
sync error detected. Suspect the X
encoder sensor or X motor.
X movement error #3.
During initialization, a motor sync
0082
error detected. Suspect the X
encoder sensor or X motor.
During an eject operation or move
operation (to a storage position),
0083
the X origin sensor could not be
detected. Suspect the X origin
sensor or X motor.
During initialization, the X origin
position could not be detected.
0084
Suspect the X origin sensor or X
motor.
X calibration error #1.
During X calibration, centering
0088
sensor OFF condition could not be
detected.
X calibration error #2.
During X calibration, centering
0089
sensor ON condition could not be
detected.
During X operation, the cartridge
008F
magazine was removed. Suspect
the magazine set sensor.
Failed to detect media in the
accessor at completion of GET
00B0
operation. Suspect the centering
sensor, X motor, or P motor.
120
Dell PowerVault TL1000 Tape Autoloader: User's Guide
Panel Indication
1. Check the accessor locking screw and
remove it if it is installed. See "Removing the
accessor locking screw" on page 27.
CHK 0080
2. Cycle the power supply and try again.
v If the problem is corrected, run Library
Verify before normal library operations
resume.
v If the problem persists, see "Contacting
CHK 0081
Dell technical support" on page 111.
CHK 0082
CHK 0083
CHK 0084
CHK 0088
CHK 0089
1. Confirm that the magazine is closed.
2. Cycle the power supply and try again.
v If the problem is corrected, run Library
CHK 008F
Verify before normal library operations
resume.
v If the problem persists, see "Contacting
Dell technical support" on page 111.
1. Confirm the media is compatible.
2. Cycle the power supply and try again.
v If the problem is corrected, run Library
CHK 00B0
Verify before normal library operations
resume.
v If the problem persists, see "Contacting
Dell technical support" on page 111.
Action Required