Measurement Modes - Kanomax 3887 Operation Manual

Handheld laser particle counter
Hide thumbs Also See for 3887:
Table of Contents

Advertisement

4. Measurement Modes

The instrument is equipped with six (6) measurement modes.
● REPEAT Mode (Suitable for repeated measurement at a same location.)
Measurement of a certain sampling period and interval can be repeated from twice to infinite
number of times. When storing the measurement data, the maximum number of measurements is
10,000 times.
● SINGLE Mode (Single measurement which stops when set sampling time has elapsed.)
Measurement stops automatically when set sampling time has elapsed.
● CONT (Continuous) Mode (Suitable for measurement of random time)
Measurement is stopped manually.
● CALC (Calculation) Mode (Processes results of repeated measurements.)
Measurement is repeated as same as the REPEAT Mode, and based on the results obtained from
the repeated measurements, average, maximum, minimum, and standard deviation are
calculated and displayed.
When storing the data in the instrument in CALC mode, only the calculation result; average,
maximum, minimum, and standard deviation will be stored, and detail data from the repeated
measurements will NOT be stored.
<Caution>
Storing the data in CALC mode requires memory 4 times larger than that of modes such as
REPEAT, SINGLE and CONT Modes. When all data is taken in CALC Mode, maximum 2,500
data records can be stored.
● ISO>4 Mode (Suitable for cleanliness assessment of ISO Class 5 to 9)
This mode is suitable for cleanliness assessment in accordance with ISO14644-1, 2 or JIS B9920.
The result is displayed by calculating the average, standard deviation and 95% UCL
automatically from the sampled data and number of sampling times.
● REMOTE Mode (Externally controlled measurement)
Measurement is controlled externally by using the software supplied with the instrument.
5

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents