Calc Mode - Kanomax 3887 Operation Manual

Handheld laser particle counter
Hide thumbs Also See for 3887:
Table of Contents

Advertisement

4-4 CALC Mode

Setup procedure is shown below.
(Character positions may not correspond to the actual screen.)
Power UP
<MENU>
1.MEASURE MODE
2.DATA PROCESSING
3.OPTIONS
Press ENTER key
< MEASURE MODE >
1.REPEAT
4.CALC
2.SINGLE
5.REMOTE
3.CONT
6.ISO>4
Press ENTER key
CALC
0.3/0.5/5.0um
LOC.021
STR:N BEEP:N
SAMPLE
10:00 10TIMES
FREE REC. 09999
Press ENTER key
CALC
30B
WAIT
0.3 0.00E+0/m3
0.5 0.00E+0/m3
00/10
5.0 0.00E+0/M3
Press ENTER key
CALC
30B
READY
0.3 0.00E+0/m3
0.5 0.00E+0/m3
00/10
5.0 0.00E+0/M3
Move the
cursor by
▲▼key.
Move the
cursor by
▲▼key.
OK
20:32
20:32
Measurement Setting Screen
LOC.:
LOC number can be used for identifying room
numbers or measurements. Setting of this item
is not mandatory.
STR:
Used when storing data in the instrument. When
storing the data, select "Y" by using the ▲▼
key. Select "N" if you do not want to store the
data.
BEEP:
Alarm goes off when alarm level is exceeded.
SAMPLE:
Sets sampling time. Setting range is from 10 sec
to 99min 59sec.
FREE REC.:
Indicates remaining data storage capacity.
Setting of every item is not mandatory.
For example, you can change only the LOC
number and press the START/STOP key, and the
display will switch to a measurement standby
screen.