Iso>4 Mode - Kanomax 3887 Operation Manual

Handheld laser particle counter
Hide thumbs Also See for 3887:
Table of Contents

Advertisement

4-6 ISO>4 Mode
Setup procedure is shown below.
(Character positions may not correspond to the actual screen.)
Power Up
<MENU>
1.MEASURE MODE
2.DATA PROCESSING
3.OPTIONS
Press ENTER key
< MEASURE MODE >
1.REPEAT
4.CALC
2.SINGLE
5.REMOTE
3.CONT
6.ISO>4
Press ENTER key
ISO>4
0.3/0.5/5.0um
LOC.021
STR:N BEEP:N
SAMPLE
01:00
INT 00:10:00
Press ENTER key
SAMPLING
SAMPLE TIME
00:01:00
TOTAL TIME
00:12:00
FREEREC.9999
Press ENTER key
ISO>4
30B
WAIT
0.5 0.00E+0/m3
5.0 0.00E+0/m3
00/02
1POINT
2TIMES
OK
OK
20:32
Measurement Setting Screen
LOC.:
LOC number can be used for identifying room
numbers or measurements. Setting of this item
is not mandatory.
STR:
Used when storing data in the instrument. When
storing the data, select "Y" by using the ▲▼
key. Select "N" if you do not want to store the
data.
BEEP:
Alarm goes off when alarm level is exceeded.
SAMPLE:
Sets sampling time. Setting range is from 10 sec
to 99min 59sec.
TIMES:
Sets number of sampling times. Setting range is
from 1 to 99, or CONT (Repeats until storage
capacity is reached. Max. 10,000 times.)
INT:
Sets duration of one measurement cycle. Setting
range is from 10sec to 99min 59sec. Minimum
time is dependent on set sampling time.
Setting of every item is not mandatory.
For example, you can change only the LOC
number and press the START/STOP key, and the
display will switch to a measurement standby
screen.
18

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents