7. Error Messages
When there is an error, the self-diagnosis function displays a symbol on the screen indicating an
error (symbol will be displayed where "■" mark is shown below).
REPEAT 30B ■
WAIT
0.3 0.00E+0/m3
0.5 0.00E+0/m3
00/02
5.0 0.00E+0/m3
Symbol
L
Laser power failure
F
Flow error
Maximum measurable
O
concentration exceeded
20:32 |
Error Status
Solution
Failure of laser light emitter.
Please contact your local distributor or our service
center.
Flow rate is exceeding the specified value
(2.83L/min ±10%).
Remove filter or tube if attached to the inlet. If the
"F" remains displayed, it may be a failure in the
flow system (including pump). Please contact your
local distributor or our service center.
Measurable concentration range is exceeded.
If the "O" symbol remains displayed even when
the measurement is performed at a cleaner
location, or by attaching the filter, please contact
your local distributor or our service center.
34