Specifications - GRAPHTEC MT100 Quick Start Manual

Hide thumbs Also See for MT100:
Table of Contents

Advertisement

12. Specifications

Standard Specifications
Item
Number of analog input
External input/output
PC interface
Internal memory devices
Sampling interval
TIME/DIV
Data save functions
(internal and USB memory)
Calculation
Display modes
Message/marker function
Data backup functions
Clock accuracy (23°C environment)
Operating environment
Withstand voltage
Insulation resistance
Power supply
Power consumption
External dimensions
Weight
Other
Interface specifications
Item
Interface types
Ethernet functions
(10BASE-T/100BASE-TX)
USB functions
Realtime data transfer speed 100 msec/10 ch maximum
Digital input specifications
Item
Input/output types
Input specifications
Alarm output specifications Output format: Relay contact output (NO/NC)
39
1 unit (10 channels)
Trigger input, Logic input 4 channels or Pulse input 4 channels, Alarm output
Ethernet (10BASE-T/100BASE-TX), USB (FullSpeed supported) provided as standard features
Internal memory: approx. 14MB
USB memory slot (FullSpeed supported) is provided as a standard feature
100, 125, 200, 250, 500ms, 1, 2, 5, 10, 20, 30sec, 1, 2, 5, 10, 20, 30min, 1 hour
1, 2, 5, 10, 20, 30 sec/DIV, 1, 2, 5, 10, 20, 30min/DIV, 1, 2, 5, 10, 12, 24, 72 hour/DIV
Save data: Measurement data, setup data, copy of data screen
Backup function: Internal to USB memory; internal to FTP server; USB memory to FTP server
* USB memory can be removed during data capture.
Types of statistical calculation: Average value, peak value, maximum value, minimum value, RMS
Calculation between channels: Addition, subtraction, multiplication, and division; calculation of humidity from wet and dry bulb temperature (Channels 1 and 2 only)
Waveform + digital (vertical and horizontal scrolling), digital, and bar graph
Function: Records message/marker at specified timing
Number of registered messages: 8
Marker type: Arbitrary message, start/stop, trigger, alarm, power failure
Setup conditions: EEPROM; Clock: lithium secondary battery
0.002% (approx. 50 seconds per month)
0 to 50°C, 5 to 85%RH
Between AC input and GND: 1 minute at 2000VAC
Between alarm terminal and GND: 1 minute at 1000VAC
Between each input channel and GND: 1 minute at 350Vp-p
Between input terminals: 1 minute at 350Vp-p
Between AC input and GND: 20M
Between alarm terminal and GND: 50M
Between each input channel and GND: 50M
AC input: 100 to 240VAC/50 to 60Hz
Terminal type: M4 screw type terminals
AC power consumption
No.
1
2
* Normal status is when LCD brightness is set to MAX.
H144 x W144 x D200mm
2.1kg
Dust-proof and water-proof
IP-65 compatible (only the front panel)
Vibration-tested conditions
Equivalent to automobile parts Type 1 Category A classification
Ethernet (10BASE-T/100BASE-TX)
USB (FullSpeed)
Web server function: Displays MT100's screen image on Web browser, operation of MT100
FTP server function: Transfers and deletes files from internal memory and USB memory
FTP client function: Supports backup of data in internal and USB memory
NTP client function: Adjusts internal clock
USB drive mode: Transfers and deletes files from internal memory
• Trigger input (1 ch) • Logic input (4 ch) or Pulse input (4 ch) • Alarm output (4 ch)
* Switch between Logic and Pulse
Maximum input voltage: 0 to +24V (single-ended ground input)
Input threshold voltage: Approx. +2.5V
Hysteresis: Approx. 0.5V (+2.5 to +3V)
Rated: 250VAC/2A
Description
or higher (500VDC)
or higher (500VDC)
or higher (500VDC)
Condition
When the LCD is ON
When the screensaver is operating
Description
Description
38VA
30VA

Advertisement

Table of Contents
loading

Table of Contents