GRAPHTEC MT100 Quick Start Manual page 22

Hide thumbs Also See for MT100:
Table of Contents

Advertisement

Input unit specifications:
Item
Number of input channels
10 ch
Input terminal type
M4 screw type terminals
Input method
Photo MOS relay scanning system
All channels isolated, balanced input
Scan speed
0.1s/10 ch maximum
Measurement
Voltage
20, 50, 100, 200, 500mV; 1, 2, 5, 10, 20, 50V; 1-5 V F.S.
ranges
Temperature
Humidity
0 to 100% (voltage 0V to 1V scaling conversion) *with B-530 (option)
Voltage
0.1% of F.S.
Measurement
accuracy*1
Temperature
(23°C 5°C
*1
When 30 minutes or more
have elapsed after power
was switched on
Sampling 1s
Filter ON (10)
GND connected
Reference contact compensation accuracy
Internal/External switching
A/D converter
16 bits (out of which 14 bits are internally acknowledged)
Temperature coefficient
Gain: 0.01% of F.S./°C
1M
Input resistance
Within 300
Allowable signal source resistance
Maximum permissible
Between +/– terminals: 60Vp-p
input voltage
Between input terminal/input terminal: 60Vp-p
Between input terminal/GND: 60Vp-p
Withstand voltage
Between input terminal/input terminal: 1 minute at 350Vp-p
Between input terminal/GND: 1 minute at 350Vp-p
Between input terminal/GND: At least 50M
Insulation resistance
At least 90dB (50/60Hz; signal source 300
Common mode rejection ratio
Noise
At least 48dB (with +/– terminals shorted)
Filter
Off, 2, 5, 10, 20, 40
Filter operation is on a moving average basis.
The average value of the set sampling count is used.
Thermocouples: K, J, E, T, R, S, B, N, W (WRe5-26)
Resistance temperature detector: Pt100, JPt100, Pt1000 (IEC751)
Thermocouple*2
Type
R/S
R: 300
S: 300
B
K
E
T
J
N
W
Reference contact compensation accuracy
*2: Thermocouple diameters T: 0.32
Resistance temperature detector
Type
Pt100
-200 to 850°C (FS=1050°C)
JPt100
-200 to 500°C (FS=700°C)
Pt1000
-200 to 500°C (FS=700°C)
5%
Description
Measurement Temperature Range
0
Ts
100
100
Ts
300
Ts
1600
Ts
1760
400
Ts
600
600
Ts
1820
200
Ts
100
100
Ts
1370
200
Ts
100
100
Ts
800
200
Ts
100
100
Ts
400
200
Ts
100
100
Ts
100
100
Ts
1100
0
Ts
1300
0
Ts
2000
, others: 0.65
Measurement Temperature Range
(at 500VDC)
or less)
Measurement Accuracy
5.2°C
3.0°C
(0.05% of rdg +2.0°C)
(0.05% of rdg +2.0°C)
3.5
(0.05% of rdg +2.0°C)
(0.05% of rdg +2.0°C)
(0.05% of rdg +1.0°C)
(0.05% of rdg +2.0°C)
(0.05% of rdg +1.0°C)
(0.1% of rdg +1.5°C)
(0.1% of rdg +0.5°C)
2.7°C
1.7°C
(0.05% of rdg +1.0°C)
(0.1% of rdg +1.0°C)
(0.1% of rdg +1.5°C)
0.5°C
Applied current
Measurement Accuracy
1mA
(0.05% of rdg +0.5°C)
1mA
(0.05% of rdg +0.5°C)
0.2mA
(0.05% of rdg +0.5°C)
40

Advertisement

Table of Contents
loading

Table of Contents