Reset The Nvp To The Default Value; Make A Custom Outlet Configuration - Fluke DSX-600 Technical Reference Handbook

Cableanalyzer
Hide thumbs Also See for DSX-600:
Table of Contents

Advertisement

DSX-600 CableAnalyzer
Technical Reference Handbook
6
Tap MEASURE.
7
On the NVP screen, use
until the measured length is the length of the cable, then tap
SAVE.

Reset the NVP to the Default Value

To reset the NVP to the default value defined by the selected
cable type, select the cable type again on the TEST SETUP screen.

Make a Custom Outlet Configuration

Custom outlet configurations start with the T568A or T568B
configurations. You can enable or disable each pair, and give the
configuration a name.
To make a custom outlet configuration
1
On the home screen, tap the test setup panel, tap a copper
test, then tap EDIT.
2
On the TEST SETUP screen, tap Outlet Configuration.
3
Tap CUSTOM, then tap MANAGE, then tap the Create panel.
4
Select the T568A or T568B configuration then select the pairs
you want enabled.
5
Enter a name for the configuration, then tap SAVE. The tester
adds an asterisk (*) to the start and end of the custom
configuration name.
6
To use the configuration, tap it on the OUTLET CONFIG
screen, then tap USE SELECTED.
186
or
to change the NVP value