Safety Messages - Emerson Rosemount Apex Reference Manual

Radar gauge
Table of Contents

Advertisement

APEX
and APEX Sentry

SAFETY MESSAGES

2-2
Radar Gauge
Procedures and instructions in this section may require special precautions to
ensure the safety of the personnel performing the operations. Information that
raises potential safety issues is indicated by a warning symbol ( ). Please
refer to the following safety messages before performing an operation
preceded by this symbol.
Explosions could result in death or serious injury:
Verify that the operating environment of the gauge is consistent with the appropriate
hazardous locations certifications.
Before connecting a HART-based communicator in an explosive atmosphere, make sure
the instruments in the loop are installed in accordance with intrinsically safe or
non-incendive field wiring practices.
Do not remove the gauge cover in explosive atmospheres when the circuit
is alive.
Failure to follow safe installation and servicing guidelines could result in death or
serious injury:
Make sure only qualified personnel perform the installation.
Use the equipment only as specified in this manual. Failure to do so may impair the
protection provided by the equipment.
Do not perform any service other than those contained in this manual unless you are
qualified.
High voltage that may be present on leads could cause electrical shock:
Avoid contact with leads and terminals.
Make sure the main power to the APEX Radar Gauge is off and the lines to any other
external power source are disconnected or not powered while wiring
the gauge.
Reference Manual
00809-0100-4731, Rev FA
April 2003

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rosemount apex sentry

Table of Contents