SETUP AND CALIBRATION, CONT.
`*$0 (*$0) - Fine Span Adjustment
NOTE! The *$0 and -`9* modes require a load of 10% of Capacity be on
the scale before adjustments can be made.
With the `*$0 prompt displayed after pressing the TARE/
prompt:
With *$0 displayed, press the TARE/
current setting 06. To skip the Fine Span Adjustment and return to the *;'$ menu, press
the TARE/
key again. To perform the Fine Span Adjustment, place a calibrated test
ENTER
weight on the scale and press the ASTERISK/
and then the TARE/
ENTER
After pressing the TARE/
weight and the annunciators will alternately flash off and on i.e. (all ON, weighing unit off,
then all OFF, weighing unit ON). Press the ASTERISK/
OR press the UNITS/LEFT ARROW key to decrease the span. Press the ASTERISK key
to return to the previous prompt or press the TARE/
to the *;'$ menu.
With the `*$0` prompt displayed after pressing the Calibration switch:
With `*$0 displayed, place a calibrated test weight on the scale and press the
TARE/
key. After pressing the TARE/
ENTER
amount of the test weight and the annunciators will alternately flash off and on (all ON,
weighing unit off, then all OFF, weighing unit ON). Press the ASTERISK/
increase the span OR press the UNITS/LEFT ARROW key to decrease the span. Press
the ASTERISK/
UP ARROW
key to exit `*$0 and return to the *;'$ menu.
-`9* - Display High Resolution Weight
With -`9* on the display, pressing the TARE/
"high resolution" mode (in 1/10 interval). Press the PRINT key to print the weight (followed by
the text TEST) via the selected printer output port enabled during setup and calibration. Press
key to return to the -`9`*`prompt. To exit the -`9`* mode, press
the TARE/
ENTER
the calibration switch or cycle power (press the ON/OFF key twice).
!66'; - Key Lock Out Function
With !66'; on the display, pressing any key will display !63 (locked) or '0!63
(unlocked) for the current key state. Pressing a locked key during normal operation will results
in a 1/2 second display !63 and the key will be ignored. To exit the !66'; function, press
the calibration switch or cycle power (press the ON/OFF key twice).
NOTE! The menu selections -`9* and !66'; can only be selected using
the calibration switch.
8200-M024-O1 Rev G y 205 Installation & Technical
32
ENTER
UP ARROW
key.
key, the display will change to show the amount of the test
ENTER
ENTER
key to return to the previous prompt or press the TARE/
ENTER
key at the last Print
ENTER
key. The display will change to show the
key (display will change to (*)
key to increase the span
UP ARROW
key to exit *$0 and return
ENTER
key, the display will change to show the
key will show the active weight in
key to
UP ARROW
ENTER