AX Carrier – Preparation
Tools
Hex socket screwdriver, 2.5 mm
Preparing the AX carrier
Before the AX carrier is installed, it may first
have to be prepared for the optics carrier to be
used (Leica M125 C, M165 C, and M205 C).
1. Check the switch position.
Leica M Series User Manual
If the switch position corresponds to the
optics carrier used, skip the following steps.
Otherwise, proceed as follows:
2. Unscrew the two screws that hold the
switch in position.
3. Pull out the switch, rotate it into the desired
position and reinsert it.
Assembly
4. Important: push the switch all the way into
the mount while reinserting the screws.
Only this ensures that the switch is posi-
tioned correctly.
27