Photo Tubes And C-Mounts - Leica M125 C User Manual

M series
Hide thumbs Also See for M125 C:
Table of Contents

Advertisement

Photo Tubes and C-mounts

Application
All Leica DFC/DMC/MC cameras are equipped
with a standardized C-mount interface. In turn,
the C-mount adapter for the respective trin-
ocular tube is connected to this interface. This
adapter creates a solid mechanical connec-
tion between the microscope and camera and
ensures optimum rendering of the microscopic
image on the image sensor of the camera.
Usually, the ideal is for the digital camera to
capture as much of the field of vision as possi-
ble, while excluding as much of the black edge
of the field of vision as possible. To do so, the
magnification factor of the C-mount adapter
must match the image format of the sensor as
closely as possible (see table).
If there is unwanted shading at the corners even
with a compatible C-mount adapter installed, it
can be corrected using the "Shading function"
of the camera software.
Leica M Series User Manual
Alternatively, you can also use a C-mount
adapter with higher magnification. This primar-
ily avoids the critical border area of the field of
vision and concentrates on the center of the
field of vision.
Cameras from third-party suppliers
In addition to Leica cameras with the standard-
ized C-mount interfaces, you can connect third-
party cameras to the microscope using a T2
bayonet adapter.
Name
MPixel Sensor Size
Pixel-Size
MC170 HD
5
CMOS-c 1/2.3" 2.35 μm
MC190 HD
10
CMOS-c 1/2.3" 1.67 μm
DMC2900
3
CMOS-c 1/2"
3.2 μm
DFC295
3
CMOS-c 1/2"
3.2 μm
DFC7000T
2.8
CCD-c
2/3"
4.54 μm
DMC4500
5
CCD-c
2/3"
3.4 μm
DFC450 C
5
CCD-c
2/3"
3.4 μm
DFC495
8
CCD-c
2/3"
2.7 μm
DFC550
12.5 CCD-c
2/3"
6.45 μm
DFC3000 G
1.3
CCD-m 1/3"
3.75 μm
DFC365 FX
1.4
CCD-m 2/3"
6.45 μm
-c: color / -m: monochrome / * recommended
Photography & Video
To do so, instead of the C-mount adapter,
simply use the corresponding SLR adapter
with T2 connection. However, these third-party
cameras are not integrated into the Leica Appli-
cation Suite and have to be operated using
the corresponding software from the camera
manufacturer.
The Leica digital cameras are detailed in a sepa-
rate user manual along with instructions for
their assembly and use.
Cooling Interface
C-Mount*
No
USB2 + HD
0.4×-0.5×
No
USB2 + HD
0.4×-0.5×
No
USB3
0.5×
No
FWb
0.5×
Yes
USB3
0.63×
No
USB3
0.63×
Yes
FWb + Sync
0.5×
Yes
FWb + Sync
0.63×
Yes
FWb + Sync
0.63×
No
USB3 + Sync 0.35×-0.4×
Yes
FWb + Sync
0.7×
Stereo 2016/V.1
74

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

M165 cM205 cM205 a

Table of Contents