13.6 Application table (recommendations/ reference values)

Prepare standard sample:
If necessary, crush sample and spread evenly in the aluminum dish.
-
Prepare special samples:
For sensitive samples or materials that are difficult to spread (e.g. mercury) a
-
glass fiber filter may be used.
Spread sample evenly on glass fiber filter and cover with a second glass fiber
-
filter.
The glass fiber filter may also be used as protection for materials that tend to
-
splash (each splash will falsify the final result).
MATERIAL
Test item
weight
g)
Dry piece of
5-8
apple
Moist piece
5-8
of apple
Butter
2-5
Mustard
2-3
Ground
2-3
coffee
Cornflakes
2-4
Yogurt
2-3
Cocoa
2-3
powder
Margarine
3-4
Milk powder
2-4
Red wine
3-5
Sunflower oil
10-14
Sugar
4-5
Milk
2-3
Flour
8-10
Cement
8-12
Paper
2-4
126
Drying
Data inquiry
temperature
interval
(º C)
100
100
138
130
106
120
110
106
138
90
100
138
138
120
130
138
106
% Moisture %
solids
(s)
10
76.5
10
7.5
15
16.3
20
76.4
5
2.8
15
9.7
15
86.5
20
0.1
20
16
15
5
15
97.4
20
0.1
15
11.9
15
88
10
12.5
15
0.8
20
6.4
MLS-BA-defsi-0713
Drying time
(min)
10-15
5-10
4.5
10
4
5-7
4.5-6.5
2
10
6
15-20
2
10
6-8
4-5
4-5
10

Hide quick links: