03 Calibration test
Here, deviation from the last adjustment is determined. This is only a check, i.e.
no values are changed.
Use the cursor (►) to select "03 Calibr. test"
Push
-button
Ensure that there are no objects on the weighing
plate.
Press
-key; the display shows the measuring
value of the adjusting weight.
Now place the adjusting weight in the center of
the weighing plate, press the
The result of the calibrating test is displayed.
Remove weight; actuate the
MLS-BA-defsi-0713
key.
-key
29.12.04
Setup
Setup
Einstellungen
Calibration check
Cear the pan
[Enter]
Einstellungen
Calibration check
Weight measuring
Einstellungen
Calibration check
Load weigh
t 50.0g
[Enter]
Einstellungen
Calibration check
Weight measuring
Calibration
Check results
Cal.: 180.1354
Act.: 180.1438
Diff.: - 0.0084
87