KERN MLS 50-3IR160 Instruction Manual page 23

Electronic moisture analyser
Table of Contents

Advertisement

.
2
Carrying out with AUTO-Temperature calibration set MLS-A01
• Temperature calibration
If you only need to measure the temperature within the MLS (only calibration),
then carry out the procedure as described under chapter 6 "Temperature
adjustment", without connecting the RS 232 interface.
Temperature adjustment
The temperature will be measured at two points and will be automatically
adjusted.
To adjust the temperature the following settings are required on the MLS-A01:
Switch on the device with the
carries out a self-test. All symbols are shown in the
display:
Note:
When calibrating the temperature, the automatic switch-off function must be
deactivated as follows:
Keep pressing the
displayed
Keep pressing the
current switch-off time is displayed.
To deactivate the function use the
to set the switch-off time to 1.
MLS-BA-defsi-0713
key. The device
key until the time is
key until A.OFF and the
or
REL
1
-
E1E2E3
CAL
alarm
alarm
alarm
keys
alarm
. . . .
%mg/l
. .
8
8
8
8
pH uS
mV mS
uMM/l
ppm ppt
P1
P2
P3
P4
P6
. . . .
-1
cm
k
%
. .
1
8
8
8
8
C
temp
time
25.2
z
°C
temp
time
. . 26
08
°14
temp
time
A.OFF
z
°1
temp
time
89