Technical Data - KERN MLS 50-3IR160 Instruction Manual

Electronic moisture analyser
Table of Contents

Advertisement

1 Technical data

Data
Radiator type
Temperature range
Maximum load
Minimum for drying
Readability (d) at weighing < 1,5 g
Readability (d) at weighing > 1,5 g
Heating methods
Reproducibility at weighing 2 g
Reproducibility at weighing 10 g
Reproducibility in weighing mode
(=standard deviation)
Recommended adjusting weight,
not included (class)
Ambient conditions
Turn-off criterion
Sample plates included
Display after drying
(*SG = Starting weight g)
Interface/RS232
Dimensions
Net weight
Electr. source
MLS-BA-defsi-0713
MLS 50-3IR160
2 Infrared radiators,
300 Watts each
50°C - 160°C
standard
fast speed
stage drying (freely selectable time / temperatura stages)
Soft
10°C ... 40°C surrounding temperature
max. 80 % air humidity (not condensing)
Automatic 1 (change of mass less than 1 mg in 20 s)
Automatic 2 (change of mass less than 1 mg in 50s)
Automatic 3 (change of mass less than 1 mg in 120s)
Automatic 4 (change of mass less than 1 mg in 180s)
Automatic 5 (change of mass less than 1 mg in 240s)
User defined
User defined 2
Test
Manual
Time-controlled (1 min – 9h 59 min)
Humidity [%] = weight loss (WL) from SG*
Dry matter [%] = residual weight (RW) from SG*
ATRO [%] [ (SG – RG) : RG ] x 100%
Residual weight [g] (RW)
Display can be changed any time
Print-out in the selected language
220-230VAC /50-60Hz or 115VAC
[ONLY to be modified by specialist]
MLS 50-3HA160 MLS 50-3HA250
2 Halogen crystal
radiators,
300 Watts each
50°C - 160°C
50 g
0.02 g
0.001 g / 0.01%
0,001 g / 0,001 %
0,5 %
0,02 %
0.001g
50g (F2)
10 (Ø 92 mm)
210 x 335 x 180 mm
6 kg
2 Halogen crystal
radiators,
300 Watts each
50°C - 250°C
0 – 100 %
100 – 0 %
0 – 999 %
Absolute value in [g]
69

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Mls 50-3ha160Mls 50-3ha250

Table of Contents