KERN MLS 50-3IR160 Instruction Manual page 22

Electronic moisture analyser
Table of Contents

Advertisement

04 Temperature calibration
We recommend to check the temperature value of the MLS50-3 from time to time.
Before the MLS50-3 should have cooled down at least 3 hours after the last heating
phase. Clamp the temperature sensor (e.g. MLS-A03) in the respective support on
the left side. Push the temperature sensor as close as possible to the thermic sensor
of the MLS.
The temperature is mesaured at two points and can be corrected at the two
temperature points.
1. Carrying out with Temperature calibration set MLB-A03
Use the cursor (►) to select
„04 Temp. calibr."
Push
-button
The temperature calibration is started.
After 14.59 Min the temp. calibration of the 1. point is
finished.
Select the keys
temperature value as necessary (e.g.24/22).
In order to take over the temperature values
press
-key, than the temp. calibration for the 2.
point is started.
After 14.59 Min the temp. calibration of the 2. point is
finished.
Select the keys
temperature value as necessary (e.g.124/120).
Press key
to take over the temperature
values.