KERN MLS 50-3IR160 Instruction Manual page 25

Electronic moisture analyser
Table of Contents

Advertisement

To adjust the temperature the following settings are required on the MLS
moisture analyzer:
When using the MLS-A01 temperature calibration set, the baud rate of the MLS must
be set to 9600 as described below:
Switch on the moisture analyzer
key.
Press the
Using the
key, select menu item "P5 RS
232"
key.
Press the
Using the cursor (►) select your settings as
"01 Baud rate"
key. The current menu item will
Press the
flash
Set the baud rate to 9600 using the arrow keys
Confirm with the
key. The balance returns to the
Press the
menu
Using the
key, select the parameter "P1
Kalibrierung"
Using the
key, select the parameter "04
Temp.kal"
key.
Press the
The temperature calibration will begin.
After 14:59 minutes, temperature calibration for the
first point is complete.
The temperature value is adjusted automatically.
MLS-BA-defsi-0713
key
Setup
P1
C
alibration
P2 GLP
P3 D
ate/Time
P4 Redout
P5 RS-232
P6
Printouts
P7
Globals
29.12.04
Setup
P 5 01 B ud rate
a
02 Parity
03 Data bits
04 Stop bits
05 Handshake
06 Print of stab
29.12.04
Setup
P1
C
alibration
P2 GLP
P3 D
ate/Time
P4 Redout
P5 RS-232
P6
Printouts
P7
Globals
29.12.04
Setup
Setup
2
9 6 0 0
0
n on e
2
8 bits
1
1 b it
n on e
0
n on e
0
91