Keithley 2260B-30-72 User Manual page 91

2260b series, 360w and 720w multi-range programmable dc power supplies
Hide thumbs Also See for 2260B-30-72:
Table of Contents

Advertisement

V-I Mode
Rising Voltage
Slew Rate
Falling Voltage
Slew Rate
88
Selects High Speed Priority or Slew Rate
Priority for CV or CC mode. The voltage or
current slew rate can only be edited if CC/CV
Slew Rate Priority is selected. The ISR indicator
will be lit for CC Slew Rate Priority and the
VSR indicator will be lit for CV Slew Rate
Priority.
Note: CC and CV Slew Rate Priority mode are
disabled when voltage/current output is set to
external control.
CC Slew Rate priority
ISR
F-03
Sets the rising voltage slew rate. Only
applicable if V-I Mode is set to CV Slew Rate
Priority.
F-04
Sets the falling voltage slew rate. Only
applicable if V-I Mode is set to CV Slew Rate
Priority.
F-05
2260B Series User Manual
CV Slew Rate priority
VSR
0 = CV high speed priority
1 = CC high speed priority
2 = CV slew rate priority
3 = CC slew rate priority
0.01V/s~60V/s (2260B-30-XX)
0.1V/s~160V/s (2260B-80-XX)
0.01V/s~60V/s (2260B-30-XX)
0.1V/s~160V/s (2260B-80-XX)