4. Use the Current knob to set the F-
03 setting.
Set F-03 to 0 (CV High Speed
Priority) or 2 (CV Slew Rate
Priority).
0 = CV High Speed Priority
F-03
2 = CV Slew Rate Priority
5. Press the Voltage knob to save the
configuration setting. ConF will be
displayed when successful.
6. If CV Slew Rate Priority was chosen as the
operating mode in steps 3~5, set F-04 (Rising
Voltage Slew Rate) and F-05 (Falling Voltage
Slew Rate) and save.
F-04 / F-05 0.1V/s~60V/s (2260B-30-XX)
0.1V/s~160V/s (2260B-80-XX)
7. Press the Function key again to exit
the configuration settings. The
function key light will turn off.
8. Use the Current knob to set the
current limit (crossover point).
Push in the knob to highlight a
digit. Turn the knob to the desired
digit to adjust. Rotate the knob to
the desired current.
2260B Series User Manual