Connection
Connection- alt.
shielding
Panel operation
102
When connecting the external voltage source to
the MIL connectors, use shielded or twisted
paired wiring.
EXT-V
+
-
2 core shielded
wire or twisted
pair
Pin16 → EXT-V (-)
Pin4 → EXT-V (+)
Wire shield → negative (-) output terminal
If the wire shield needs to be grounded at the
voltage source (EXT-V), then the shield cannot
also be grounded at the negative (-) terminal
output of the 2260B power supply. This would
short the output.
EXT-V
+
-
2 core shielded
wire or twisted
pair
Pin16 → EXT-V(-)
Pin4 → EXT-V(+)
Wire shield → EXT-V ground (GND)
1. Connect the external voltage according to the
connection diagrams above.
2260B Series User Manual
Power
supply
4
Analog
connector
16
Output
Terminal
Power
supply
4
Analog
connector
16
Output
Terminal