Keithley 2260B-30-72 User Manual page 62

2260b series, 360w and 720w multi-range programmable dc power supplies
Hide thumbs Also See for 2260B-30-72:
Table of Contents

Advertisement

Output Voltage/
Output Current
Slew Rate
Only the master unit's current or voltage
slew rate settings apply.
The settable current slew rate is double or
triple that of a single unit, depending on the
number of units that are used in parallel.
For example:
The 2260B-30-36 has a maximum settable
current slew rate of 72A/S. When two 2260-
30-36 units are used in parallel, the
maximum settable slew rate setting is
144A/S.
Internal Resistance
Only the master unit's internal resistance
settings apply.
For 2 units in parallel, the internal resistance
is actually half of the setting value.
For 3 units in parallel, the internal resistance
is actually a third of the setting value.
Bleeder Control
The master unit is used to control the bleeder
settings. The bleeder resistors in all the slave
units are always turned off when in parallel
mode.
Model
2260B-30-36
2260B-80-13
2260B-30-72
2260B-80-27
Single unit
2 units
30V
30V
36A
72A
80V
80V
13.5A
27A
30V
30V
72A
144A
80V
80V
27A
54A
OPERATION
3 units
30V
108A
80V
40.5A
30V
216A
80V
81A