Keithley 2260B-30-72 User Manual page 76

2260b series, 360w and 720w multi-range programmable dc power supplies
Hide thumbs Also See for 2260B-30-72:
Table of Contents

Advertisement

Cycle Step Start: Determines the starting step for the Cycle
Repeat field. Range: 1~999. The Cycle Step Start field must be less
than the Cycle Step End field.
Cycle Step End: Determines the ending step for the Cycle Repeat
field. Range: 2~1000. The Cycle Step End field must be greater
than the Cycle Step Start field.
Line 2:
Step Header: This field contains the "Step" header text. Required.
Do not remove from this location.
Time(sec) Header: This field contains the "Time(sec)" header text.
Required. Do not remove or alter from this location.
Voltage (V) Header: This field contains the "Voltage (V)" header
text. Required. Do not remove or alter from this location.
Current (A) Header: This field contains the "Current (A)" header
text. Required. Do not remove or alter from this location.
Line 3+:
Step Number: The Step Number field numbers each step in the
test function. It has a range of 1~1000. All steps must be
numbered in numerical order. No step number can be "skipped"
between the first and last step.
Time(sec): Designates the duration time of the corresponding
step in seconds. It has a range of 0.05 seconds to 20 days. It has a
step resolution of 0.05 seconds (50ms). A setting of 0 will skip the
step.
Voltage (V): Designates the voltage of the corresponding steps in
volts. MIN or MAX can be used to set the voltage to minimum or
maximum, respectively.
Current (A): Designates the current of the corresponding steps in
amps. MIN or MAX can be used to set the current to minimum or
maximum, respectively.
OPERATION
73

Advertisement

Table of Contents
loading

This manual is also suitable for:

2260b-30-362260b-80-132260b-80-27

Table of Contents