Instruction Manual
IM-106-880C, OI
7.8.1
Support Resource Block Errors
Resource Block
Out of Service
•
Power Up
•
Block Configuration Error
•
Simulation Active
•
Transducer Block
•
Input Failure
•
Simulation Active
Other Error
•
7.9
Transducer Block
The Transducer Block was designed to provide the information necessary to interface Hazardous
Area OCX 8800 to the Fieldbus.
7.9.1
Transducer Block Parameters
Table 7-4
Fieldbus specifications the description can be found.
Support Resource Block
Errors
Set Whenever the resource block actual mode is OOS.
Set whenever the FF card powers up.
Configuration Error is used to indicate that you had selected an item in FEATURES_SEL
or CYCLE_SEL that was not set in FEATURES or CYCLE_TYPE, respectively.
Set whenever the Fieldbus Simulate Switch is set to ON at the Fieldbus card or software
simulate option enabled.
Out of Service
–
Set whenever the transducer block actual mode is OOS.
Set whenever there is a communication error between the Fieldbus card and the OCX.
–
Set whenever the Fieldbus Simulate Switch is set to ON at the Fieldbus card or
software simulate option enabled.
Set whenever XD_ERROR is non-zero.
gives a description of all parameters, or gives the location of the
Section 7: Foundation
Fieldbus
April 2017
10