Emerson Rosemount OCX8800 Instruction Manual page 123

Oxygen and combustibles transmitter, hazardous location probe with integral electronics
Hide thumbs Also See for Rosemount OCX8800:
Table of Contents

Advertisement

Instruction Manual
IM-106-880C, OI
Alerts
Alarms
Sensor Heater
Transducer
Malfunction
Block
1) O
Heater
2
Failure
2) Comb
Heater Failure
3) SB Heater
Failure
4) O
Heater
2
Ramp Rate
Exceeded
5) Comb
Heater Ramp
Rate Exceeded
6) SB Heater
Ramp Rate
Exceeded
7) Heater Relay
Failed
Sensor Heater
Transducer
Over Temp-
Block:
erature
1) O
Cell
2
Temp. Very
High
2) CombTemp.
Very High
3) SB Temp
Very High
Transducer
Sensor Heater
Block:
Temp-
1) O
Cell
erature
2
Temp. Low
Variance
2) O
Cell
2
Temp. High
3) CombTemp.
Low
4) CombTemp.
High
5) SB Temp.
Low
6) SB Temp.
High
Transducer
Calibration
Block:
Error
1) O
2
Calibration
Failed
2) Comb
Calibration
Failed
3) Calibration
Warning
Mapping of PWA
What does
the alert
AMS
Tab
indicate?
T/C
This alert
Heater
indicates that
Alerts
no measurable
heat energy is
being detected check
at the oxygen
sensor or that
the heater
temperature is
rising too fast.
TC
This alert
Heater
indicates a very
Alerts
high heater
temperature;
temperature is
rising too fast.
T/C
This alert
Heater
indicates a
Alerts
sensor heater
temperature
that is too high
or too low.
Calib-
This alert
ration
indicates that
Alerts
the slope and
constant
values
determined
from the
calibration did
not fall within
an acceptable
range.
Recommended
Action
Description
Check heater
Mechanical or thermal stress may eventually
circuit for lose or
cause the sensor heater to fail. The resistance
broken
of a properly functioning cell heater will
connections,
measure less than 100 ohms. A failed heater
will generally measure as an open circuit.
thermocouple
Diagnosis must be done at the analyzer. Refer
wiring, test or
to Section 8, Overview page 165 for details.
replace the
heater.
The Heater Ramp Rate Exceeded problem is
usually caused by the inability of the device
to limit power to the heater. This could be
caused by a shorted triac component on the
power supply in the electronics stack.
Diagnosis must be done at the analyzer. Refer
to Section 8, Overview page 165 for details.
Check heater
A heater over-temperature/out of control
wiring, check
problem would generally be caused by the
thermocouple
inability of the device to limit power to the
wiring or replace
heater. This could be caused by a shorted
the electronics
triac on the power supply in the electronics
stack.
stack.
Allow instrument
Cell temperature control may become erratic
several minutes to
for the following reasons:
reach proper
1) Temperature is settling during startup.
temperature or
2) Large variations in process temperature or
check power
flow.
supply.
3) Fluctuations or noise in the power supplied
to the instrument.
Check the
Make sure the oxygen concentrations of the
calibration gas
calibration gases match the concentration
supplies and
values in the device. If the calibration has
connections.
been performed correctly this alarm may
indicate that the sensor requires
replacement. Refer to Section 8, Overview
page 165 for details.