To Use I2C Triggering - Agilent Technologies MSO6012A User Manual

6000 series
Hide thumbs Also See for MSO6012A:
Table of Contents

Advertisement

2
To use I
C triggering
N O T E
6000 Series Oscilloscope User's Guide
2
For I
C decode setup see
2
An I
C (Inter-IC bus) trigger setup consists of connecting the
oscilloscope to the serial data (SDA) line and the serial clock
(SCL) line, then triggering on a stop/start condition, a restart, a
missing acknowledge, an EEPROM data read, or on a read/write
frame with a specific device address and data value.
1 Press the More key in the Trigger section of the front panel,
rotate the Entry knob until I
softkey, then press the Settings softkey to display the I
trigger menu.
Trigger on:
Signals
condition
menu
Triggering the Oscilloscope
page
266.
2
C is displayed in the Trigger
Currently selected Clock
2
I
C trigger
Data
Address
value
2
C
Trigger level
or Data channel
or threshold
Data 2
Return to
value
previous menu
4
161

Advertisement

Table of Contents
loading

Table of Contents