Agilent Technologies MSO6012A User Manual page 285

6000 series
Hide thumbs Also See for MSO6012A:
Table of Contents

Advertisement

6
Displaying Data
284
b Press the Source softkey, and use the Entry knob to select
the channel probing the LIN signal.
c Repeatedly press and release the Baud softkey to specify
the LIN signal baud rate.
d Repeatedly press and release the Smpl Pt softkey to select
the sample point. This controls the point within the bit's
time where the bit value is captured.
One Bit
Sample Point
e Repeatedly press and release the Standard softkey to select
the LIN standard (LIN 1.3 or LIN 2.0) of the measured
signal. For LIN 1.2 signals, use the LIN 1.3 setting. The LIN
1.3 setting assumes the signal follows the "Table of Valid
ID Values" as shown in section A.2 of the LIN Specification
dated December 12, 2002. If your signal does not comply
with the table, use the LIN 2.0 setting.
f Repeatedly press and release the Sync Break softkey to
define the minimum number of clocks that define a Sync
Break.
4 Make sure the trigger or threshold levels are set to the middle
of the LIN signal.
For analog channels, turn the Trigger Level knob.
For digital channels, press the D15 Thru D0 key and the
Thresholds softkey to access the threshold level setting
softkeys.
60%
70%
80%
6000 Series Oscilloscope User's Guide