Integrate - Agilent Technologies MSO6012A User Manual

6000 series
Hide thumbs Also See for MSO6012A:
Table of Contents

Advertisement

Integrate

6000 Series Oscilloscope User's Guide
∫ dt (integrate) calculates the integral of the selected source.
You can use integrate to calculate the energy of a pulse in
volt-seconds or measure the area under a waveform.
∫ dt plots the integral of the source using the "Trapezoidal
Rule". The equation is:
n
+ ∆t
I
= c
y
n
o
i
i=0
Where
I = integrated waveform
∆t = point-to-point time difference
y = channel 1, 2, or function 1 + 2, 1 – 2, and 1 * 2 data
points
c
= arbitrary constant
o
i = data point index
In Delayed sweep horizontal mode, the ∫ dt function does not
display in the delayed portion of the display.
1 Press the Math key, press the ∫ dt softkey, then press the
Source, Scale, or Offset sofkeys if you want to change the
source, scaling, or offset for the integrate function.
Source — selects the source for ∫ dt. The source can be
any analog channel, or math functions 1 + 2, 1 – 2,
and 1 * 2.
Scale — lets you set your own vertical scale factors for
∫ dt expressed in unit-seconds/division, where units can
be V (Volts), A (Amps), or W (Watts). Units are set in
the channel Probe menu. Press the Scale softkey, then
turn the Entry knob to rescale ∫ dt.
Making Measurements
5
213

Advertisement

Table of Contents
loading

Table of Contents