Post-Maintenance Checkout; Preparing The Ex3000 For Storage; Table 7-1 Ex3000 Post-Maintenance Checkout - Vetland EX3000 Operator's Manual

Veterinary anesthesia system
Table of Contents

Advertisement

Post-Maintenance Checkout

After performing maintenance on the EX3000, perform a post-maintenance checkout as described in
Table 7-1.
Verify
No pressure leaks
Proper O
flush
2
button operation
Proper operation of
Inspiration and
Expiration Valves
Proper operation of
the Pop-Off Valve

Preparing the EX3000 for Storage

1. Turn OFF (0) the EX3000 Anesthesia System.
2. Disconnect the main power cable from the facility AC power supply.
3. Turn off the oxygen supply valve and disconnect the facility oxygen supply line (if used). Remove
and safely store E-size stand-mounted oxygen cylinders (if used).
4. Remove and discard disposable breathing tubes and bags in accordance with local regulations.
5. Empty the CO
absorber canister and discard the absorbent material in accordance with the
2
manufacturer's MSDS and local regulations.
6. Drain both anesthetic agent reservoirs as specified earlier in this section.
7. Clean and disinfect all surfaces and connections as specified earlier in this section.
8. Protect the system from dust and contamination with a plastic cover or other suitable means.
Maintenance

Table 7-1 EX3000 Post-Maintenance Checkout

Test Method/ Acceptance Criteria
Perform a System Pressure Leak Test in accordance with
Troubleshooting.
Verify that the button does not stick.
Visually verify that the valves move freely and do not stick.
Verify that the valve easily opens and closes and does not stick.
Section 6 –
7-7