Facility Requirements; Table 1-1 Facility Requirements - Vetland EX3000 Operator's Manual

Veterinary anesthesia system
Table of Contents

Advertisement

Facility Requirements

Table 1-1 provides the facility requirements necessary to ensure reliable operation and safety of the
EX3000 Anesthesia System.
Characteristic
Working Area
Power
Oxygen Supply
Pneumatic Fitting
Environment
Note: If you have any questions about facility requirements, please contact Vetland or your
authorized distributor.
Introduction

Table 1-1 Facility Requirements

Large enough to accommodate the anesthesia system. See Figure 1-2.
95 - 240 VAC, 50/60 Hz
Note: An AC power fail alarm sounds with both audible and visual indicators if
power supply drops below 95 VAC. The alarm is self-resetting.
Delivery Options:
Oxygen supply hose from the facility oxygen outlet
Type E cylinder mounted to the anesthesia machine via an optional
manifold and yoke assembly
Type H cylinder strapped to facility wall or a cylinder cart
Regulated Inlet Pressure: 50–60 psi (345–380 kPa)
Note: A low O
pressure alarm sounds with both audible and visual indicators
2
if supply pressure drops below approximately 45 psig. Alarm is self-resetting.
Flow Rate: 20 liters per minute at 55 psi (379 kPa)
Oxygen: Medical Grade
Diameter Indexed Safety System (DISS) for Oxygen
Storage Temperature: 41 to 122 F (5 to 50 C)
Operating Temperature: 59 to 95 F (15 to 35 C)
Relative Humidity: 5 to 90% (non-condensing)
Altitude: 1609 meters (5278 feet) maximum
Specification
1-3